کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1877298 1042064 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Performance of a whole-body counter with five high-purity germanium detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Performance of a whole-body counter with five high-purity germanium detectors
چکیده انگلیسی

A germanium (Ge) detector system has been installed in a whole-body counting facility at NIRS (Japan). The system employs five detectors that have relative efficiencies of more than 80%. A basic performance of the Ge detectors was tested and compared with that of the former NaI(Tl) detector system. Although the Ge detector system has advantages of improved radionuclide identification capability and MDA values, it also has disadvantages of high operation costs and demanding maintenance such as supply of liquid nitrogen.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 64, Issue 3, March 2006, Pages 386–389