• The ion-beam-induced charge collection technique was used for charge collection study.
• Experimental results are in good agreement with data of Monte-Carlo simulations.
• The detector is capable of measuring α-particle activity of contaminated soil samples.
Circular ion-implanted silicon detector of α-particles with a large, 5-cm2, sensitive area has been developed. An advantage of the detector is that the detector surface is easily cleanable with chemicals. The hardened surface of the detector shows no signs of deterioration of the spectroscopic and electrical characteristics upon repeated cleaning. The energy resolution along the diameters of the detector was (1.0±0.1)% for the 5.486-MeV α-particles. Detailed tests of the charge collection efficiency and uniformity of the detector entrance window were also performed with a 5.5-MeV He2+ microbeam.
Journal: Applied Radiation and Isotopes - Volume 92, September 2014, Pages 96–101