A new model calculation of the peak efficiency for HPGe detectors used in assays of radioactive waste drums
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
A new model calculation of the peak efficiency for HPGe detectors used in assays of radioactive waste drums
چکیده انگلیسی
In this paper we present a new semi-empirical model calculation of the peak efficiency for unshielded HPGe detectors based on the virtual point detector and the attenuation factor concepts. The validity of the model calculation was checked by comparison with Monte Carlo efficiency values and experimental efficiencies determined for a HPGe detector type GEM 25P4 using a calibration drum. The discrepancy between experimental and calculated efficiencies is smaller than 10% in the energy range 122-1408 keV.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 68, Issues 7–8, July–August 2010, Pages 1418-1422