کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1878229 | 1042289 | 2009 | 5 صفحه PDF | دانلود رایگان |

k0 Instrumental Neutron Activation Analysis (k0 INAA) was evaluated at the University of Missouri Research Reactor (MURR). The MURR, unlike many other research reactors that employ k0 INAA, is refueled on a weekly basis. To determine if the neutron spectrum is stable enough for routine k0 INAA analysis, the neutron spectrum parameters α and f were measured over a 9 month period. The average values of f and α measured using the Cd ratio multi monitor technique were 57.4±4.5 and 0.039±0.012, respectively. It was determined that k0 INAA could be used at the MURR by employing the average value of α and f. However, significant measurement bias could occur for elements with high Q0 values if the true spectral parameters deviate by more than 10% from the average value.
Journal: Applied Radiation and Isotopes - Volume 67, Issue 6, June 2009, Pages 1084–1088