Development of a transfer standard for the measurement of low Rn-222 activity concentration in air
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Development of a transfer standard for the measurement of low Rn-222 activity concentration in air
چکیده انگلیسی
The transfer standard consists of a multi-wire impulse ionization chamber, electronic measurement equipment and the corresponding software. The multi-wire impulse ionization chamber is composed of two parallel printed boards. Between the boards, 478 vertical electrode wires are soldered in to include a volume of 10 L. In the dependency of the energy, the α-particles of Rn-222 and the decay products are detected in the active volume of the multi-wire impulse ionization chamber. By means of the measurement software an α-energy spectrum is produced. Via different algorithms, the activity concentration of the Rn-222 reference atmosphere is determined.