کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1878686 | 1532110 | 2014 | 4 صفحه PDF | دانلود رایگان |
• The installation of the new Compton Suppression System at NED-NTUA is presented.
• Simultaneous collection of both suppressed and unsuppressed spectrum.
• Peak-to-Compton improvement by a factor of ~5.
• Peak Suppression Factors range from 1 to 0.066, depending on energy and isotope.
• MDA reduction by 13–50%.
This paper presents the Compton Suppression System, recently installed at the Nuclear Engineering Department of NTUA. The system consists of an XtRa Ge detector coupled with a NaI(Tl) guard detector. The electronic set-up allows for the simultaneous collection of both the suppressed and the unsuppressed spectra. System performance is investigated using certified point and volume sources. Parameters such as Peak Suppression Factors, peak-to-Compton ratios and minimum detectable activity for specific radionuclides are determined.
Journal: Applied Radiation and Isotopes - Volume 87, May 2014, Pages 361–364