کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1878749 | 1532115 | 2013 | 5 صفحه PDF | دانلود رایگان |

• Length-compensated internal proportional counting is discussed.
• We described the development of low-background unequal-length proportional counters.
• We described the development of unequal-length OFHC proportional counters.
• A gas mixing and transfer system is described.
• An uncertainty analysis, following the ISO GUM guidelines, is presented.
Pacific Northwest National Laboratory (PNNL) is developing a capability to measure the absolute activity concentration of gaseous radionuclides using length-compensated proportional-counting. This capability will enable the validation of low-level calibration standards for use in PNNL's new shallow underground laboratory. Two sets of unequal length proportional counters have been fabricated; one set has been fabricated using ultra-low background (ULB) electroformed copper and a second set fabricated from Oxygen-Free High-Conductivity Copper (OFHC).
Journal: Applied Radiation and Isotopes - Volume 81, November 2013, Pages 179–183