کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1878829 1532117 2013 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
COMS eye plaque brachytherapy dosimetric sensitivity to source photon energy and seed design
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
COMS eye plaque brachytherapy dosimetric sensitivity to source photon energy and seed design
چکیده انگلیسی
This study explores the influence of source photon energy on eye plaque brachytherapy dose distributions for a 16 mm COMS plaque filled with 103Pd, 125I, or 131Cs sources or monoenergetic photon emissions ranging from 12 keV to 100 keV. Dose distributions were similarly created for all permutations of three common brachytherapy seed designs. Within this range, sources with average energy ≤22 keV may reduce dose to the opposite eye wall by more than a factor of 2 while maintaining tolerable proximal sclera doses when prescribing to depths of 9 mm or less. Current commercially-available brachytherapy sources can exhibit up to 15% relative dosimetric sensitivity to seed design at regions within the eye.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 79, September 2013, Pages 62-66
نویسندگان
, ,