1878963 | 1042529 | 2013 | 6 صفحه PDF | دانلود رایگان |

The main aim of this paper was to compare different methods of preparing water samples for tritium analysis by ultra-low-level background liquid scintillation counter, Quantulus 1220. Three methods of sample preparation for low-level tritium measurement have been implemented in the Nuclear Physics Laboratory in Novi Sad: electrolytic enrichment, direct method without electrolytic enrichment and sample Oxidizer 307 method. The examined fresh water samples were rainfall collected during 6 months and water from a stream in the Vinča nuclear research center collected over 3 months. The obtained results using these three methods showed satisfying agreement. The appropriate measuring time by LSC for each sample prepared according to different methods has been determined.
► 3 methods of water sample preparation for low-level tritium measurement have been compared.
► All prepared samples were measured by LSC Quantulus 1220.
► The most novel is that Sample Oxidizer is used as one method for sample preparation.
► Paper presents MDA values for each presented method.
► For environmental monitoring all three methods give comparable results.
Journal: Applied Radiation and Isotopes - Volume 71, Issue 1, January 2013, Pages 51–56