کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1878977 | 1532097 | 2015 | 4 صفحه PDF | دانلود رایگان |

• Thermoluminescent dosimetry of LiF:Mg,Ti.
• Comparation between doses calculated by the LiF:Mg,Ti and doses provided by the planning system.
• Use of PMMA phantom for IMRT photon beam.
• Optically stimulated luminescence (OSL) dosimetry.
The principle of IMRT is to treat a patient from a number of different directions (or continuous arcs) with beams of nonuniform fluences, which have been optimized to deliver a high dose to the target volume and an acceptably low dose to the surrounding normal structures (Khan, 2010). This study intends to provide information to the physicist regarding the application of different dosimeters type, phantoms and analysis technique for Intensity Modulated Radiation Therapy (IMRT) dose distributions evaluation. The measures were performed using dosimeters of LiF:Mg,Ti and Al2O3:C evaluated by techniques of thermoluminescent (TL) and Optically Stimulated Luminescence (OSL). A polymethylmethacrylate (PMMA) phantom with five cavities, two principal target volumes considered like tumours to be treated and other three cavities to measure the scattered radiation dose was developed to carried out the measures.
Journal: Applied Radiation and Isotopes - Volume 100, June 2015, Pages 7–10