A commercial solution of 65Zn was standardized by the sum peak-method using a planar HPGe detector. The activity results were compared with measurements made with a well type 4πγ ionization chamber, which is traceable to BIPM.RI (II)-K2.Zn-65 key-comparison performed in 2002. The sum-peak value was 42.79 kBq/g and the ionization chamber value was 42.74 kBq/g both at the reference date. The uncertainty obtained in the sum peak standardization was 0.25% (k=1), and in the ionization chamber was 0.85% (k=1).The results showed that sum-peak method can be used in 65Zn standardization and this method is easier, simpler and more practical than others methods.
► A solution of 65Zn was standardized by sum peak-method using a planar HPGe detector.
► The nuclide was standardized using the coincidence with 8 keV X-ray and 1115 keV g-ray.
► To measure the 1115 and 1123 g-rays, a device voltage divider was introduced in the pre-amplifier output.
► This method can be seen as an alternative in relation to other methods.
Journal: Applied Radiation and Isotopes - Volume 70, Issue 9, September 2012, Pages 2087–2090