Comparison of experimental and calculated calibration coefficients for a high sensitivity ionization chamber
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Comparison of experimental and calculated calibration coefficients for a high sensitivity ionization chamber
چکیده انگلیسی

The response of a Vacutec 70129 ionization chamber was calculated using the PENELOPE-2008 Monte Carlo code and compared to experimental data. The filling gas mixture composition and its pressure have been determined using IC simulated response adjustment to experimental results. The Monte Carlo simulation revealed a physical effect in the detector response to photons due to the presence of xenon in the chamber. A very good agreement is found between calculated and experimental calibration coefficients for 17 radionuclides.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 70, Issue 9, September 2012, Pages 2232–2236