کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1879787 | 1042593 | 2006 | 10 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Some results of a simulated test for administration of activity in nuclear medicine
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
This paper describes the results obtained using a simulated test for administration of activity in nuclear medicine between 2002 and 2004. Measurements in the radionuclide calibrator are made during the different stages of the procedure. The test attempts to obtain supplementary information on the quality of the measurement, with the aim of evaluating in a more complete way the accuracy of the administered activity value compared with the prescribed one. The participants' performance has been assessed by means of a statistical analysis of the reported data. Dependences between several attributes of the simulated administration tests results are discussed. Specifically, the proportion of satisfactory results in the 2003-2004 period was found to be higher than in 2002. It reveals an improvement of the activity administration in the Cuban nuclear medicine departments since 2003.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 64, Issue 4, April 2006, Pages 475-484
Journal: Applied Radiation and Isotopes - Volume 64, Issue 4, April 2006, Pages 475-484
نویسندگان
P. Oropesa, A.T. Hernández, R.A. Serra, C. Varela, M.J. Woods,