کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1879876 1532131 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Application of PENELOPE code to the efficiency calibration of coaxial germanium detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Application of PENELOPE code to the efficiency calibration of coaxial germanium detectors
چکیده انگلیسی
The code PENELOPE is applied here to the efficiency calibration exercise described in the EUROMET project 428 [Lépy et al., 2001. Appl. Radiat. Isot. 55, 493]. We evaluated the peak efficiencies for a coaxial HPGe detector in the range 60-2000 keV, for point sources located at various distances from the detector and for a cylindrical box containing two different matrixes: a low-density silica and a hydrochloric acid solution. After an optimization of several detector parameters, the results were in good agreement with the experimental values.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 64, Issues 10–11, October–November 2006, Pages 1319-1322