کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1879888 | 1532131 | 2006 | 5 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Comparison of strontium-89 solution sources in UK hospitals, 2003
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله
چکیده انگلیسی
Concerns have been raised within the nuclear medicine field about the accuracy of measurements of 89Sr. The 85Sr impurity present in the 89Sr solution has a significant effect on the response of radionuclide calibrators used in hospitals. A comparison was conducted between the National Physical Laboratory (NPL) and the UK hospital physics community. Only 58% of the results were within 5% of the NPL value. Work at NPL has resulted in correction factors for the NPL secondary standard radionuclide calibrator and for Capintec calibrators, which will produce a significant improvement in performance.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 64, Issues 10â11, OctoberâNovember 2006, Pages 1375-1379
Journal: Applied Radiation and Isotopes - Volume 64, Issues 10â11, OctoberâNovember 2006, Pages 1375-1379
نویسندگان
M.J. Woods, M.I. Baker, D.K. Tyler, K. Chari,