The performance of a prototype device designed to evaluate general quality parameters of X-ray equipment
ترجمه فارسی عنوان
عملکرد یک دستگاه نمونه اولیه طراحی شده برای ارزیابی کیفیت کلی پارامترهای تجهیزات اشعه ایکس
کلمات کلیدی
سیستم تجزیه و تحلیل مایع چند منظوره، آشکارساز حالت جامد، اتاق یونیزه، تست های کنترل کیفیت
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


• An instrument was developed to evaluate the performance of radiological equipment.
• The prototype containing solid detector was compared with commercial instruments.
• The prototype device shows comparable reproducibility and accuracy with commercial instruments.
• The prototype can be used for evaluate kVp, exposure time, and HVL using a single X-ray exposure.

The performance of radiological equipment can be assessed using non-invasive methods and portable instruments that can analyze an X-ray beam with just one exposure. These instruments use either an ionization chamber or a state solid detector (SSD) to evaluate X-ray beam parameters. In Brazil, no such instruments are currently being manufactured; consequently, these instruments come at a higher cost to users due to importation taxes. Additionally, quality control tests are time consuming and impose a high workload on the X-ray tubes when evaluating their performance parameters. The assessment of some parameters, such as the half-value layer (HVL), requires several exposures; however, this can be reduced by using a SSD that requires only a single exposure. One such SSD uses photodiodes designed for high X-ray sensitivity without the use of scintillation crystals. This sensitivity allows one electron-hole pair to be created per 3.63 eV of incident energy, resulting in extremely high and stable quantum efficiencies. These silicon photodiodes operate by absorbing photons and generating a flow of current that is proportional to the incident power. The aim of this study was to show the response of the solid sensor PIN RD100A detector in a multifunctional X-ray analysis system that is designed to evaluate the average peak voltage (kVp), exposure time, and HVL of radiological equipment. For this purpose, a prototype board that uses four SSDs was developed to measure kVp, exposure time, and HVL using a single exposure. The reproducibility and accuracy of the results were compared to that of different X-ray beam analysis instruments. The kVp reproducibility and accuracy results were 2% and 3%, respectively; the exposure time reproducibility and accuracy results were 2% and 1%, respectively; and the HVL accuracy was ±2%. The prototype's methodology was able to calculate these parameters with appropriate reproducibility and accuracy. Therefore, the prototype can be considered a multifunctional instrument that can appropriately evaluate the performance of radiological equipment.