کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1883054 1533504 2016 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
INGAS: Iranian Noble Gas Analyzing System for radioxenon measurement
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
INGAS: Iranian Noble Gas Analyzing System for radioxenon measurement
چکیده انگلیسی


• A new radioxenon detection system based on coincidence has been developed.
• The detection system consisted of a well-typed NaI(Tl) and a plastic scintillator.
• The MDA for 133Xe was determined to be 1.240 mBq during 1 day measurement time.
• The performance has been evaluated by injection of beta–gamma emitter sources.

In this article, Iranian Noble Gas Analyzing System (INGAS) will be introduced. This system is based on beta–gamma coincidence technique and consists of a well-type NaI(Tl) as gamma or X radiation detector and a cylindrical plastic scintillator to detect beta or conversion electron. Standard NIM modules were utilized to detect coincidence events of detectors. Both the beta and gamma detectors were appropriately calibrated. The efficiency curve of gamma detector for volume geometry was obtained by comparing the results of gamma point sources measurements and simulations of GATE V7.0 Monte Carlo code. The performance of detection system was checked by injection of 222Rn and 131mXe gaseous source in the detection cell. The minimum detectable activity of the system for 133Xe is 1.240±0.024 mBq for 24 h measurement time.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 120, March 2016, Pages 26–32
نویسندگان
, , , , ,