Overbeaming and overlapping of volume-scan CT with tube current modulation in a 320-detector row CT scanner
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Overbeaming and overlapping of volume-scan CT with tube current modulation in a 320-detector row CT scanner
چکیده انگلیسی
The purpose of this study was to evaluate the performance of volume scan tube current modulation (VS-ATCM) with adaptive iterative dose reduction 3D (AIDR3D) technique in abdomen CT examinations. We scanned an elliptical cone-shaped phantom utilizing AIDR3D technique combined with VS-ATCM mode in a 320-detector row CT scanner. The image noise distributions with conventional filtered back-projction (FBP) technique and those with AIDR3D technique were compared. The radiation dose profile and tube current time product (mAs) in three noise levels of VS-ATCM modes were compared. The radiation beam profiles of five preset scan lengths were measured using Gafchromic film strips to assess the effects of overbeaming and everlapping. The results indicated that the image noises with AIDR3D technique was 13-74% lower than those in FBP technique. The mAs distributions can be a prediction for various abdominal sizes when undergoing a VS-ATCM mode scan. Patients can receive the radiation dose of overbeaming and overlapping during the VS-ATCM mode scans.