Evaluation of radiation dose to pediatric patients during certain special procedures
ترجمه فارسی عنوان
ارزیابی دوز تابش در بیماران کودکان در طی مراحل خاص
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


• Pediatric radiation dose has been evaluated for three of the most common fluoroscopic procedures.
• Radiation doses were measured using calibrated TLD GR200A.
• Pediatric patients of concern and ESAK doses showed large variations.
• The patient doses in this study are within the reported studies suggesting that the pediatric patients are adequately protected.

This study was intended to measure pediatric entrance surface air kerma (ESAK) and effective dose during micturating cystourethrography (MCU), intravenous urography (IVU) and barium studies (barium meal, enema, and swallow) and to propose a local diagnostic reference level (DRL). ESAK was measured for patients using calibrated thermoluminescent dosimeters (TLDs, GR200A). Effective doses (E) were calculated using the National Radiological Protection Board (NRPB) software. A total of 236 special pediatric procedures were investigated. 21.7% of the sample comprised barium procedures, 18.6% were MCU procedures while 59.5% of the sample were IVU procedures. The mean ESAK measurements (mGy) were 2.1±0.8, 3.0±23 and 1.2±0.2 for barium meal, enema and swallow in the same order. The mean patient dose for IVU procedures was 12.4±8.7 mGy per procedure and the mean patient dose per MCU procedure was 5.8±7 mGy. Local DRLs were proposed for all procedures. The patient doses in this study are within the reported values, suggesting that pediatric patients are adequately protected.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 104, November 2014, Pages 267–271
نویسندگان
, , , , ,