کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1883431 | 1533520 | 2014 | 4 صفحه PDF | دانلود رایگان |
• Determination of the focused X-ray beam parameters is presented.
• Beam intensity is derived using a method based on detecting characteristic X-rays.
• Dose rates are derived from the established fluence rates.
This paper deals with estimating the dose rates achieved in materials irradiated by X-rays from an X-ray tube equipped with polycapillary focusing optics. The proposed method consists of determining the main parameters of the X-ray beam, e.g. the shape of the spectrum, the beam intensity, and the beam profile. All these parameters were obtained with a single spectrometric detector, and they were then employed in evaluating the dose rate. The maximum achievable dose rates in iron and water were determined analytically, and a numeric calculation using the Monte Carlo method provided the depth distribution of the energy deposited in these two materials.
Journal: Radiation Physics and Chemistry - Volume 104, November 2014, Pages 436–439