کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1885885 | 1533515 | 2015 | 4 صفحه PDF | دانلود رایگان |
• The pulse reconstruction method was utilized to perform neutron-gamma discrimination (NGD) with NE213 scintillator.
• A digital storage oscilloscope with an 8-bit digital-to-analog convertor was responsible for receiving the scintillator anode pulses.
• The NGD results on an Am–Be combined neutron-gamma source represents the feasibility of the proposed method.
In this research, based on pulse reconstruction method, digital storage oscilloscopes with an 8-bit digital-to-analog convertor was used to successfully perform the neutron-gamma discrimination with NE213 (or its equivalent BC501A) scintillator anode pulses at minimum discrimination bias value of 95 keVee (or keV electron equivalent). Also, a 100 mCi 241Am–Be source and micro-Curie gamma ray sources (137Cs and 22Na) were used for the system calibration and discrimination studies. The results confirm the feasibility and simplicity of the proposed method.
Journal: Radiation Physics and Chemistry - Volume 109, April 2015, Pages 13–16