کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1886106 | 1533526 | 2014 | 9 صفحه PDF | دانلود رایگان |
• We investigate dosimetric characteristics of commercial Ge-doped optical fibres.
• We develop audit methodology for measuring absorbed dose under reference conditions.
• Ge-doped optical fibres offer promising thermoluminescence (TL) properties.
• Audit methodology has been developed with an expanded uncertainty of 4.22%.
Ge-doped optical fibres offer promising thermoluminescence (TL) properties together with small physical size and modest cost. Their use as dosimeters for postal radiotherapy dose audits of megavoltage photon beams has been investigated. Key dosimetric characteristics including reproducibility, linearity, dose rate, temperature and angular dependence have been established. A methodology of measuring absorbed dose under reference conditions was developed. The Ge-doped optical fibres offer linearity between TL yield and dose, with a reproducibility of better than 5%, following repeated measurements (n=5) for doses from 5 cGy to 1000 cGy. The fibres also offer dose rate, angular and temperature independence, while an energy-dependent response of 7% was found over the energy range 6 MV to 15 MV (TPR20,10 of 0.660, 0.723 and 0.774 for 6, 10 and 15 MV respectively). The audit methodology has been developed with an expanded uncertainty of 4.22% at 95% confidence interval for the photon beams studied.
Journal: Radiation Physics and Chemistry - Volume 98, May 2014, Pages 33–41