ObjectivesAn LC-MS/MS method was developed for simultaneous quantitation of tacrolimus, sirolimus and everolimus in whole blood, and compared to HPLC-UV and immunoassay methods.Design and methodsBlood (0.1 mL) was analysed following solid-phase extraction and chromatographic resolution using a C18 column (45 °C) and mobile phase of methanol/40 mM ammonium acetate/glacial acetic acid (83/17/0.1) at 200 μL/min, with positive electrospray ionisation and multiple reaction monitoring.ResultsIntra- and inter-day imprecision and inaccuracy were ≤ 12.2% over a 1.5–40 μg/L calibration range. An external quality assurance programme confirmed acceptable inaccuracy and imprecision of the LC-MS/MS method, but highlighted problems with immunoassay quantitation, particularly for everolimus, showing a > 30% bias in FPIA everolimus concentrations measured in pooled patient samples versus spiked drug-free whole blood.ConclusionsLC-MS/MS provides significant accuracy and precision advantages compared to HPLC and immunoassays. Discrepancies in everolimus concentrations measured by the Seradyn FPIA immunoassay require further investigation.
Journal: Clinical Biochemistry - Volume 44, Issues 2–3, February 2011, Pages 231–236