ObjectivesTo confirm the analytical performance of the Dimension Vista LOCI troponin I assay (cTnI).Design and methodLimit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ) with a 10% coefficient of variation (CV), linearity, precision, method comparison, and 99th percentile upper reference limits (URL) were analyzed. Endogenous analytes and rheumatoid factor (RF) were tested for assay interference.ResultsThe 99th percentile was 0.022 µg/L (CV = 14%) and the LoQ was 0.036 µg/L. The ratio of 10% CV concentration to 99th percentile was 1.63. Linearity extended from 0 to 44.36 µg/L. The method comparison equation was Dimension® Vista™ = 0.94 (Dimension® RxL) + 0.00 µg/L with bias at low levels. No interference was detected.ConclusionsThis study shows acceptable performance characteristics of the LOCI cTnI assay on Dimension® Vista™ to diagnosis and risk stratification of patients with acute coronary syndrome symptoms.
Journal: Clinical Biochemistry - Volume 43, Issue 12, August 2010, Pages 998–1002