کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1970591 1538899 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Assessing agreement between point of care and laboratory results for lipid testing from a clinical perspective
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی زیست شیمی
پیش نمایش صفحه اول مقاله
Assessing agreement between point of care and laboratory results for lipid testing from a clinical perspective
چکیده انگلیسی

Objectives:Investigate agreement between lipid pathology results from point-of-care testing (PoCT) devices and laboratories.Design and methods:Agreement was assessed using the Bland–Altman method.Results:Mean difference (limits of agreement) were: − 0.28 mmol/L (− 1.04, 0.48) for total cholesterol, − 0.09 mmol/L, (− 0.55, 0.36) for HDL-C. Median difference (nonparametric limits of agreement) were 0.07 mmol/L, (− 0.40, 3.04) for triglycerides.Conclusions:The clinical acceptability of the variation between lipid PoCT and laboratory test results is debatable but our work provides baseline data for further research.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Biochemistry - Volume 43, Issues 4–5, March 2010, Pages 515–518
نویسندگان
, , , , , , , , , , , ,