کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1970781 1059820 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A two-center international evaluation of the Immulite 2000 automated serum gastrin assay
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی زیست شیمی
پیش نمایش صفحه اول مقاله
A two-center international evaluation of the Immulite 2000 automated serum gastrin assay
چکیده انگلیسی

Objectives:To assess the performance of the Immulite 2000 serum gastrin assay.Design and methods:Two-center validation and comparison with two manual gastrin assays.Results:Serum, serum separator, EDTA and heparin tube results differed <20%. Samples showed <20% degradation for <6 h at room temperature, 3-day-refrigerated, and underwent 3 freeze–thaw cycles. Imprecision was <8% between 20 and 824 ng/L (10–393 pmol/L). Spike recovery was 88.5–106%, and recovery on dilution was 62–135%. There were no interferences from gastrin 1–14, pentagastrin, and cholecystokinin (CCK) 1–33 and CCK 26–33. The fasting reference range was <100 ng/L (<48 pmol/L). Regression slopes to the manual assays were 0.9 and 0.97, with comparable clinical performance.Conclusions:The Immulite 2000 assay is precise, accurate, and fast and compares well with established gastrin assays.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Biochemistry - Volume 39, Issue 4, April 2006, Pages 387–390
نویسندگان
, , , , , ,