کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2159685 1090864 2009 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی تحقیقات سرطان
پیش نمایش صفحه اول مقاله
Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study
چکیده انگلیسی

IntroductionOur aim was to implement standards for quality assurance of IGRT devices used in our department and to compare their performances with that of a CT simulator.Materials and methodsWe investigated image quality parameters for three devices over a period of 16 months. A multislice CT was used as a benchmark and results related to noise, spatial resolution, low contrast visibility (LCV) and uniformity were compared with a cone beam CT (CBCT) at a linac and simulator.ResultsAll devices performed well in terms of LCV and, in fact, exceeded vendor specifications. MTF was comparable between CT and linac CBCT. Integral nonuniformity was, on average, 0.002 for the CT and 0.006 for the linac CBCT. Uniformity, LCV and MTF varied depending on the protocols used for the linac CBCT. Contrast-to-noise ratio was an average of 51% higher for the CT than for the linac and simulator CBCT. No significant time trend was observed and tolerance limits were implemented.DiscussionReasonable differences in image quality between CT and CBCT were observed. Further research and development are necessary to increase image quality of commercially available CBCT devices in order for them to serve the needs for adaptive and/or online planning.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiotherapy and Oncology - Volume 93, Issue 1, October 2009, Pages 1–7
نویسندگان
, , , , , ,