کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2160089 1548898 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
High precision transponder localization using a novel electromagnetic positioning system in patients with localized prostate cancer
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی تحقیقات سرطان
پیش نمایش صفحه اول مقاله
High precision transponder localization using a novel electromagnetic positioning system in patients with localized prostate cancer
چکیده انگلیسی

Background and purposeThe Micropos 4DRT system is being developed to provide accurate, precise, objective, and continuous target localization during radiotherapy. This study involves the first in vivo use of the system, aiming to evaluate the localization accuracy of this electromagnetic positioning technique compared with radiographic localization and to assess its real-time tracking ability.Material and methodsAn active positioning marker was inserted in the prostatic urethra of 10 patients scheduled to receive radiotherapy for localized prostate cancer. A receiving sensor plate (antennae system) was placed at a known position in the treatment tabletop. Initial in vivo system calibrations were performed in three subjects. Ten additional patients were then enrolled in a study arm that compared radiographic transponder location to radiotransponder location simultaneously acquired by the Micropos 4DRT system. Frontal and side radiographs were taken with the radiopaque transponder located at three different positions within the prostatic urethra.ResultsThe transponder insertions were all successful and without complications. Comparison of the transponder location as per the Micropos 4DRT system with the radiographic transponder localization showed an average (±SD) absolute and relative 3D difference of 2.7 ± 1.2 and 1.7 ± 1.0 mm, respectively. Continuous transponder tracking capability was also demonstrated.ConclusionsElectromagnetic positioning using the Micropos transponder system is feasible in vivo. Evaluation of this novel non-ionizing localization system, in this study using a transponder positioned in the prostatic urethra, indicates transponder localization accuracy to isocenter comparable with X-ray localization of a radiopaque marker.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiotherapy and Oncology - Volume 90, Issue 3, March 2009, Pages 307–311
نویسندگان
, , , , , , ,