کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2161187 1090907 2006 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Motion and shape change when using an endorectal balloon during prostate radiation therapy
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی تحقیقات سرطان
پیش نمایش صفحه اول مقاله
Motion and shape change when using an endorectal balloon during prostate radiation therapy
چکیده انگلیسی

PurposeTo investigate motion and shape change when using an endorectal balloon (ERB) in patients receiving radiotherapy for prostate cancer.MethodsIn nine patients treated for prostate cancer using an ERB, the anterior wall of the ERB was contoured on right lateral images taken immediately before irradiation, and on left lateral images taken immediately after irradiation. Changes in the contours were used to calculate inter-fraction shape change and inter-imaging motion and shape change. Inter-imaging motion describes changes that occur after the right lateral image is taken that are seen in the left lateral image.ResultsEighty-six percent of all inter-imaging shifts of the anterior wall of the ERB were in the posterior direction (mean: 1.8 mm, 1 SD: 1.8 mm, maximum posterior shift: 2.8–7.2 mm). The inter-fraction shape change (1 SD) of the anterior wall was equivalent to a change in the angle of the balloon of 2.5–5.7°, with a range of 8–20°, depending on the patient. Inter-imaging shape changes were similar in size.ConclusionsThe inter-imaging motion and shape changes may be explained by the patient relaxing some time after insertion of the ERB, indicating that it could be reduced by a waiting period after insertion before irradiation. Development of image-guided localization strategies should consider intra-fraction motion and also inter- and intra-fraction shape change.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiotherapy and Oncology - Volume 81, Issue 2, November 2006, Pages 184–189
نویسندگان
, , , ,