Optical quality and intraocular scattering assessed with a double-pass system in eyes with contact lens induced corneal swelling
ترجمه فارسی عنوان
کیفیت نوری و پراکندگی داخل چشم به وسیله یک سیستم دو نفره در چشم ها با تورم قرنیه ناشی از لنزهای تماسی بررسی شده است
کلمات کلیدی
لنز طبی، یک شب، تماس با تورم، کیفیت نوری، پراکندگی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
چکیده انگلیسی

PurposeTo evaluate the impact of contact lens (CL)-induced corneal swelling on the optical quality of the eye by means of the double-pass technique.MethodsMeasurements of 6 healthy subjects were obtained in 5 visits over 1 week, at baseline and after sleeping with 4 different CLs of +0.50 D, +2.00 D, +5.00 D and +8.00 D (Acuvue2), randomly fitted on 4 different days. The control eye wore no CL. Corneal pachymetry and optical quality of the eye (OQAS, Visiometrics) were measured once at baseline and at three interval times in the follow-up visits: immediately after CL removal, and 1 and 2 h after CL removal. Optical quality was evaluated by means of the Strehl ratio and OQAS values at 100%, 20% and 9% contrasts. Intraocular scattering was evaluated with the objective scatter index (OSI).ResultsMean overnight swelling was 5.98 ± 4.29% in CL-eyes versus 0.30 ± 0.78% in control eyes (p < 0.01). Corneal swelling was maximal immediately after CL removal and decreased with time (p < 0.01). A significant worsening in all optical quality parameters and a significant increase of the OSI were found in eyes with corneal swelling (p < 0.05). Two hours after CL removal there were no statistically significant differences (p > 0.05) between CL-eyes and control eyes in any of the measured parameters.ConclusionsCorneal swelling has a significant impact on the optical quality of the eye and on intraocular scattering as assessed with the double-pass technique.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Contact Lens and Anterior Eye - Volume 37, Issue 4, August 2014, Pages 278–284
نویسندگان
, , , , , ,