• A diamond detector has been characterized for use as neutron yield monitor of a portable 14 MeV neutron generator.
• The system will be used for the 14 MeV calibration of JET neutron detector.
• The results and the performances of the monitor are very satisfactory in term of accuracy and reliability.
A new Deuterium-Tritium (DT) campaign is planned at JET. An accurate calibration for the 14 MeV neutron yield monitors is necessary. In order to perform the calibration a 14 MeV Neutron Generator with suitable intensity (∼108 n/s) will be used. Due to the intensity change during the Neutron Generator lifetime it would be necessary to monitor continuously the neutron emission intensity during the calibration using a compact detector attached to it. A high quality diamond detector has been chosen as one of the monitors. This detector has been fully characterized at the 14 MeV Frascati Neutron Generator facility. The characterization procedure and the resulting 14 MeV neutron response of the detector are described in this paper together with the obtained uncertainties.
Journal: Fusion Engineering and Design - Volume 106, May 2016, Pages 93–98