Dose reduction with iterative reconstruction in multi-detector CT: What is the impact on deformation of circular structures in phantom study?
ترجمه فارسی عنوان
کاهش دوز با بازسازی تکراری در CT چند آشکارساز: تأثیر بر تغییر شکل ساختارهای دایره ای در مطالعه فانتوم چیست؟
کلمات کلیدی
توموگرافی کامپیوتری؛ بازسازی فضایی؛ کیفیت تصویر؛ تغییر ساختار دایره ای؛ طیف قدرت نویز CNR، نسبت به نسبت نویز؛ CTDI، شاخص توزيع کادميوم؛ FBP، پروژکتور پشتی فیلتر شده؛ FOV، زمینه دید؛ IR، reco تکراری
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی انفورماتیک سلامت
چکیده انگلیسی

ObjectivesTo evaluate the distortion of circular structures induced by the increased image noise related to dose reduction and to assess the effect of iterative reconstruction (IR).MethodsMDCT acquisitions were performed with 120 kVp for 200/100/60/40/20 mAs with 100%/50%/30%/20%/10% of dose. Raw data were reconstructed by filtered back projection (FBP) and with two IR strengths. Image quality indices referred to water and acrylic were measured on a quality image phantom. Areas, perimeters, circularity were measured on the circular inserts of 4.8, 7.9 and 11.1 mm on a morphological phantom.ResultsDose reduction resulted in increased image noise and in decreased signal to noise ratio and contrast to noise ratio. IR improved these indices for the same dose without affecting the signal (number CT) and spatial resolution (modulation transfer function). The values of area, perimeter and circularity were altered compared to the actual value and the inserts were visually deformed with the dose reduction. IR improved these three parameters. Image quality indices, areas, perimeters and circularity of inserts were similar between the acquisition at 100% of the dose in FBP, 50% in strength-3 and 30% in strength-5 with different curves of noise power spectrum.ConclusionIR associated to 70% of dose reduction modifies the images smooth (NPS) but maintains adequate image quality indices without causing distortions of circular structures.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Diagnostic and Interventional Imaging - Volume 97, Issue 2, February 2016, Pages 187–196
نویسندگان
, , , , , , ,