BackgroundCervical spine radiography is a commonly employed examination for degenerative disease and trauma in the cervical spine. Traditionally, the anteroposterior projection is undertaken with the use of an anti-scatter grid. Some practitioners appear to have rejected this practice in favour of a non-grid technique, possibly because of the dose saving it affords. It is necessary to determine if image quality in the cervical spine is significantly degraded and whether the omission of the grid is justified.MethodUsing a slaughtered lamb neck as a model of the human neck triplicate radiographs were obtained using a non-grid, a stationary grid and a moving grid technique. Entrance surface dose and dose area product was measured for these techniques. Image quality in terms of contrast, sharpness and overall acceptability was evaluated by 9 independent and blinded observers.ResultsA significant reduction in measured dose was observed when the non-grid technique was compared to stationary or moving grid techniques. A statistically significant reduction in image contrast, sharpness and acceptability was also seen in the non-grid compared to grid techniques.ConclusionThese results show evidence of significantly greater image quality in the presence of either a moving or stationary grid in the lamb model. As such they support the continued use of scatter rejection methods such as the anti-scatter grid in AP radiography of the human cervical spine, to optimise radiographic image quality in this critical structure.
Journal: Radiography - Volume 17, Issue 2, May 2011, Pages 139–144