Objective and subjective comparison of the visibility of three echogenic needles and a nonechogenic needle on older ultrasound devices
ترجمه فارسی عنوان
مقایسه هدف و ذهن از دید سه سوزن اکسیژنیک و سوزن غیرخوجی در دستگاه های سونوگرافی
کلمات کلیدی
سوزن اکسیژنیک دید سوزنی، بلوک عصبی تحت هدایت سونوگرافی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
چکیده انگلیسی

ObjectiveThis study evaluated the visibility of echogenic needles with older ultrasound devices in an in vitro phantom study.MethodsWe compared three echogenic needles from B. Braun (BB), Unisis (UN), and Hakko (HK) with a nonechogenic needle. Each needle was inserted into an ultrasound phantom 10 times at 30° and 45° with the bevels up. The captured images of the needle and background contrast were digitally analyzed, and the median of 10 insertions for each angle was calculated to determine objective needle visibility. Needle images were also shown to 12 anesthesiologists to evaluate subjective visibility on a five-point Likert scale.ResultsThe shafts of all the echogenic needles were significantly more objectively visible than the nonechogenic needles. Subjective visibilities of the BB and UN needles were significantly higher than that of the nonechogenic needle. Therefore, the BB and UN needles were judged to have more than fair subjective visibility. However, subjective visibility of the HK needle was consistently and significantly lower than that of the BB and UN echogenic needles. At 45°, the HK needle had nearly the same poor visibility as the nonechogenic needle.ConclusionThe results of our study indicate that the BB and UN needles are more visible than nonechogenic needles in an ultrasound phantom, even on older devices.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Acta Anaesthesiologica Taiwanica - Volume 53, Issue 1, March 2015, Pages 1–6
نویسندگان
, , , , ,