کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2768593 1151266 2014 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Programación errónea de una bomba de TCI. Caso SENSAR del trimestre
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
پیش نمایش صفحه اول مقاله
Programación errónea de una bomba de TCI. Caso SENSAR del trimestre
چکیده انگلیسی
We report the case of a patient who underwent surgical aortic valve replacement. During general anaesthesia maintenance, the patient received a remifentanyl infusion via a target controlled infusion (TCI) system. The infusion pump that was prepared to deliver the infusion showed malfunction at the beginning of the surgery, so it was quickly replaced with a second pump. After a few minutes into the surgery, the patient presented with hypotension refractory to treatment. The remifentanyl syringe also emptied faster than expected. On reviewing the TCI pump, it was found that it was erroneously programmed for propofol instead of remifentanyl, thus the patient had received a very high dose of remifentanyl that was probably the cause of the haemodynamic disturbances. The incident was an error in equipment use, facilitated by hurry, lack of checking of the equipment prior to its use, and the complex and unclear design of the devices' screens. After analysis of this incident, all TCI pumps were reviewed, and all the programs for infrequently used drugs were deleted. Furthermore, 2 pumps were selected for exclusive use in the cardiac surgery theatre, one with propofol-only programming, and the other with remifentanyl-only programming, both clearly marked and situated in fixed places in that theatre.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Revista Española de Anestesiología y Reanimación - Volume 61, Issue 8, October 2014, Pages e27-e30
نویسندگان
,