کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2777382 1567976 2015 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of the analytical performance of the novel NS-Prime system and examination of temperature stability of fecal transferrin compared with fecal hemoglobin as biomarkers in a colon cancer screening program
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی بیوشیمی بالینی
پیش نمایش صفحه اول مقاله
Evaluation of the analytical performance of the novel NS-Prime system and examination of temperature stability of fecal transferrin compared with fecal hemoglobin as biomarkers in a colon cancer screening program
چکیده انگلیسی


• Performance of the NS-Prime for fecal transferrin and hemoglobin is acceptable.
• Performance of the NS-Prime for fecal transferrin and hemoglobin is acceptable.
• 95th percentile for fecal transferrin in a reference population was 4.9 µg/g feces.
• Temperature stability of transferrin and hemoglobin in feces is similar.

ObjectiveTo examine the analytical aspects of fecal transferrin (Tf) and hemoglobin (Hb) measured on the NS-Prime analyzer for use in a colon cancer screening program.Designs and methodsMethod evaluation and temperature stability studies for fecal Tf and Hb were completed. A method comparison was carried out against the NS-Plus system using samples collected from 254 screening program participants. A further 200 samples were analyzed to help determine suitable reference limits for fecal Tf using these systems.ResultsThe assay for fecal Tf showed acceptable linearity, precision, and recovery, and showed minimal carryover with low potential for impact by the prozone effect. The 95th percentile for fecal Tf obtained for the reference population was 4.9 µg/g feces. The collection device sufficiently maintained fecal Tf and Hb stability for at least 7 days at room temperature, 4 °C, and −20 °C. Fecal Tf and Hb were most stable at 4 °C and −20 °C, but showed considerable loss (20–40%) of both proteins at 37 °C within the first 7 days. Mixing small amounts of blood into diluted fecal samples maintained at 37 °C for various time periods showed >50% loss of both proteins within 1 h of incubation.ConclusionsThe NS-Prime analyzer showed acceptable performance for fecal Tf and Hb. These studies suggest that use of both Tf and Hb together as biomarkers will result in higher positivity rates, but this may not be attributed to greater stability of Tf over Hb in human feces.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Practical Laboratory Medicine - Volume 2, 1 August 2015, Pages 29–36
نویسندگان
, , , , , , , , , ,