کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2915237 1575559 2006 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Depopulated Bovine Ureteric Xenograft for Complex Haemodialysis Vascular Access
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Depopulated Bovine Ureteric Xenograft for Complex Haemodialysis Vascular Access
چکیده انگلیسی

ObjectivesTo assess performance of a decellularised bovine ureter vascular graft for haemodialysis in patients for whom conventional access was not possible.MethodsThe SynerGraft® Vascular Graft Model 100 (SGVG 100) is a bovine ureter modified by a tissue-engineering depopulation technology and uniquely it is not chemically cross-linked. SGVG 100 was implanted in patients with a failed fistula or vascular access grafts. Graft patency was the primary outcome; secondary outcomes included adverse events and associated treatments.Results25 SGVG 100 were implanted in 23 patients; mean age was 59±14 years. Mean follow-up was 370 days. The mean time to occlusion (19 events) was 215±141 days with patency re-established in 14 of 18 surgical interventions. Thirty angioplasties were performed on 14 SGVG 100 for luminal/anastomotic stenosis. Two grafts demonstrated areas of dilation; however, both grafts continue to be usable at last reported follow-up (930 and 602 days) with no further changes in graft size. Primary patency, assisted primary patency, secondary patency, and freedom from infection were 29, 45, 81, and 95% at 1 year, respectively.ConclusionsThis report demonstrates SGVG 100 is a stable vascular access conduit, providing a suitable graft alternative when autologous vein is not available.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Vascular and Endovascular Surgery - Volume 31, Issue 2, February 2006, Pages 181–186
نویسندگان
, , , ,