Impella LP 5.0 (Abiomed) is a relatively new technology that has been used over the last few years as a ventricular assist device to provide full circulatory support for patients presenting with cardiogenic shock. There is limited evidence available in literature with regard to insertion approaches, safety, complications and long-term clinical outcomes. Complications that have been reported are device malfunction, pump displacement, intravascular haemolysis, intra-ventricular thrombosis, functional mitral stenosis due to displacement and local vascular complications at vascular access site. We report here the first case with acute mitral regurgitation due to chordal rupture and flail mitral valve leaflet as a result of Impella device displacement, a new complication. Review of literature is also reported.
Journal: Heart, Lung and Circulation - Volume 23, Issue 3, March 2014, Pages e100–e104