چکیده انگلیسی

BackgroundIn Boston Scientific dual-chamber devices, the RYTHMIQ algorithm aims to minimize right ventricular pacing.ObjectiveWe evaluated the performance of this algorithm determining (1) the appropriateness of the switch from the AAI(R) mode with backup VVI pacing to the DDD(R) mode in case of suspected loss of atrioventricular (AV) conduction and (2) the rate of recorded pacemaker-mediated tachycardia (PMT) when AV hysteresis searches for restored AV conduction.MethodsIn this multicenter study, we included 157 patients with a Boston Scientific dual-chamber device (40 pacemakers and 117 implantable cardioverter-defibrillators) without permanent AV conduction disorder and with the RYTHMIQ algorithm activated. We reviewed the last 10 remote monitoring–transmitted RYTHMIQ and PMT episodes.ResultsWe analyzed 1266 episodes of switch in 142 patients (90%): 207 (16%) were appropriate and corresponded to loss of AV conduction, and 1059 (84%) were inappropriate, of which 701 (66%) were related to compensatory pause (premature atrial contraction, 7%; premature ventricular contraction, 597 (56%); or both, 27 (3%)) or to a premature ventricular contraction falling in the post-atrial pacing ventricular refractory period interval (219, 21%) and 94 (10%) were related to pacemaker dysfunction. One hundred fifty-four PMT episodes were diagnosed in 27 patients (17%). In 85 (69%) of correctly diagnosed episodes, the onset of PMT was directly related to the algorithm-related prolongation of the AV delay, promoting AV dissociation and retrograde conduction.ConclusionThis study highlights some of the limitations of the RYTHMIQ algorithm: high rate of inappropriate switch and high rate of induction of PMT. This may have clinical implications in terms of selection of patients and may suggest required changes in the algorithm architecture.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Heart Rhythm - Volume 13, Issue 6, June 2016, Pages 1266–1273
نویسندگان
, , , , , , , , , ,