کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2922262 1175841 2013 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans
چکیده انگلیسی

BackgroundCatheter-tissue contact force (CF) determines radiofrequency (RF) ablation lesion size. Impedance changes during RF delivery are used as surrogate markers for CF. The relationship between impedance and real-time CF in humans remains unknown.ObjectivesTo determine whether impedance changes have predictive value for real-time CF during catheter ablation of atrial arrhythmias.MethodsReal-time CF, force-time integral, and impedance were measured in 2265 RF lesions for atrial fibrillation or flutter in 34 patients. Operators were blinded to CF measurements. Impedance preablation, at 5-second intervals for 30 seconds after the RF onset, maximal impedance fall and time to impedance plateau during RF were correlated with CF. Average CF was divided into low (≤20 g), intermediate (21–60 g), and high (>60 g) categories.ResultsPreablation impedance poorly correlated with preablation CF (R = .07). Maximal impedance fall modestly correlated with average CF and force-time integral (R = .32 and .37, respectively). There was a large degree of overlap in impedance fall between different CF categories. A maximal impedance fall of 10 Ω could predict average CF of >20 g, with a sensitivity and specificity of 71% and 53% and a positive and negative predictive value of 51% and 49%, respectively. Impedance fall was only able to differentiate between different CF categories ≥15 seconds after the RF onset. Higher CFs moderately correlated with delayed plateau in impedance (R = .41).ConclusionsImpedance measurements (both baseline and impedance fall) are, at best, moderately efficacious as surrogate markers for predicting real-time catheter-tissue CF. These findings highlight the importance of real-time CF measurements, rather than impedance changes to optimize ablation efficacy.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Heart Rhythm - Volume 10, Issue 7, July 2013, Pages 962–969
نویسندگان
, , , , , , , , , , , , , , ,