کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2941728 1177083 2008 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Reduced Risk of Restenosis in Small Vessels and Reduced Risk of Myocardial Infarction in Long Lesions With the New Thin-Strut TAXUS Liberté Stent : 1-Year Results From the TAXUS ATLAS Program
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Reduced Risk of Restenosis in Small Vessels and Reduced Risk of Myocardial Infarction in Long Lesions With the New Thin-Strut TAXUS Liberté Stent : 1-Year Results From the TAXUS ATLAS Program
چکیده انگلیسی

ObjectivesThe TAXUS ATLAS Small Vessel (SV) and Long Lesion (LL) multicenter studies compared the performance of the thin-strut (0.0038 inch) TAXUS Liberté 2.25-mm stent (Boston Scientific; Natick, Massachusetts) and the TAXUS Liberté 38-mm long stent (Boston Scientific; Natick, Massachusetts) with the earlier paclitaxel-eluting TAXUS Express (Boston Scientific) stent that has identical polymer, drug dosage, and release kinetics but different stent geometry and thicker struts (0.0052 inch).BackgroundThe TAXUS Liberté stent was designed with thinner and more even strut spacing to provide more uniform drug distribution, as well as increased flexibility and conformability. Clinical benefits of the new stent design have not been evaluated.MethodsThe TAXUS ATLAS SV and LL studies are nonrandomized studies comparing outcomes of the TAXUS Liberté 2.25 mm (N = 261) and TAXUS Liberté 38 mm (N = 150) stents to TAXUS Express historical control groups derived from the TAXUS IV and V trials. Inclusion/exclusion criteria for TAXUS Express and Liberté groups were similar in both studies.ResultsEach study met its primary end point of noninferiority of 9-month in-segment diameter stenosis. Furthermore, TAXUS Liberté 2.25 mm, when compared with TAXUS Express, significantly reduced the rate of both 9-month angiographic restenosis (18.5% vs. 32.7%, p = 0.0219) and 12-month target lesion revascularization (6.1% vs. 16.9%, p = 0.0039). In addition, TAXUS Liberté 38 mm significantly reduced the risk of 12-month myocardial infarction compared with TAXUS Express (1.4% vs. 6.5%, p = 0.0246).ConclusionsThe thinner-strut TAXUS Liberté stent improved outcomes compared with the earlier TAXUS Express stent in both SVs and LLs (A Study of the TAXUS Liberté Stent for the Treatment of de Novo Coronary Artery Lesions in Small Vessels; NCT00371748; A Study of the TAXUS Liberté Stent for the Treatment of Long De Novo Coronary Artery Lesions; NCT00371475).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: JACC: Cardiovascular Interventions - Volume 1, Issue 6, December 2008, Pages 699–709