کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2942576 1177135 2015 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Identification and Management of Pump Thrombus in the HeartWare Left Ventricular Assist Device System : A Novel Approach Using Log File Analysis
ترجمه فارسی عنوان
شناسایی و مدیریت ترومبوز پمپ در دستگاه کمک رسانی بطن چپ قلب: روش روان با استفاده از تجزیه و تحلیل فایل ورودی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
چکیده انگلیسی

ObjectivesThe study sought to characterize patterns in the HeartWare (HeartWare Inc., Framingham, Massachusetts) ventricular assist device (HVAD) log files associated with successful medical treatment of device thrombosis.BackgroundDevice thrombosis is a serious adverse event for mechanical circulatory support devices and is often preceded by increased power consumption. Log files of the pump power are easily accessible on the bedside monitor of HVAD patients and may allow early diagnosis of device thrombosis. Furthermore, analysis of the log files may be able to predict the success rate of thrombolysis or the need for pump exchange.MethodsThe log files of 15 ADVANCE trial patients (algorithm derivation cohort) with 16 pump thrombus events treated with tissue plasminogen activator (tPA) were assessed for changes in the absolute and rate of increase in power consumption. Successful thrombolysis was defined as a clinical resolution of pump thrombus including normalization of power consumption and improvement in biochemical markers of hemolysis. Significant differences in log file patterns between successful and unsuccessful thrombolysis treatments were verified in 43 patients with 53 pump thrombus events implanted outside of clinical trials (validation cohort).ResultsThe overall success rate of tPA therapy was 57%. Successful treatments had significantly lower measures of percent of expected power (130.9% vs. 196.1%, p = 0.016) and rate of increase in power (0.61 vs. 2.87, p < 0.0001). Medical therapy was successful in 77.7% of the algorithm development cohort and 81.3% of the validation cohort when the rate of power increase and percent of expected power values were <1.25% and 200%, respectively.ConclusionsLog file parameters can potentially predict the likelihood of successful tPA treatments and if validated prospectively, could substantially alter the approach to thrombus management.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: JACC: Heart Failure - Volume 3, Issue 11, November 2015, Pages 849–856
نویسندگان
, , , , , , , , , , , , ,