Electrophysiological validation of total atrial conduction time measurement by tissue doppler echocardiography according to age and sex in healthy adults
ترجمه فارسی عنوان
اعتبار سنجی الکتروفیزیولوژیک اندازه گیری زمان کل هدایت دهلیزی توسط اکوکاردیوگرافی داپلر بافت با توجه به سن و جنس در بزرگسالان سالم
کلمات کلیدی
اعتبار سنجی، مجموع زمان هدایت دهلیزی، تصویربرداری داپلر بافت، مطالعه الکتروفیزیولوژیک
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
چکیده انگلیسی

BackgroundWe sought to validate total atrial conduction time (TACT) measurement via tissue Doppler imaging (TDI) by comparing the electrophysiological study (EPS) measurements of healthy subjects, according to age and sex.MethodsEighty patients with normal EPS results were included. TACT was measured by EPS and TDI. For validation, the results of TDI were compared with those of EPS. TACT was assessed by measuring the time interval between the beginning of the P-wave on the surface ECG, and the peak A-wave on TDI from the left atrial lateral wall, just over the mitral annulus. Electrophysiological TACT was defined as the time from the high right atrial electrogram to the distal coronary sinus atrial electrogram around the left lateral portion of the mitral ring.ResultsEPS and TDI measurements of the TACT were significantly and positively correlated among men and women in 20–30 years (p=0.008, r=0.412; p>0.001, r=0.706, respectively), and those in the 30–40 years group (p=0.001, r=0.649; p=0.001, r=0.696). In contrast, EPS and TDI measurements of TACT were not significantly different among men and women in the 20–30 years and those in the 30–40 years group (p>0.05, for both). On univariate regression analyses, TACT was independently associated with age (β=0.342, =0.001).ConclusionsWhen assessed according to the age and sex of healthy participants, TDI and EPS measurements during TACT assessments were similar and correlated with each other. The measurement of TACT via TDI may be used accurately and confidently than the measurement via EPS in healthy individuals.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Arrhythmia - Volume 32, Issue 2, April 2016, Pages 127–132
نویسندگان
, , , , , , ,