کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
2957620 | 1178183 | 2015 | 7 صفحه PDF | دانلود رایگان |

BackgroundElectrode positioning is important for the operation of implantable loop recorders (ILRs). This study aimed to investigate optimal electrode positions for the implantation of ILRs.MethodsFifteen male subjects participated in the study and R wave amplitude data were collected from 15 areas in the left anterior chest area, with 4 variations of electrode angles and body positions.ResultsThe estimated minimum R wave amplitude values were greater than 0.3 mV (manufacturer-recommended value) when electrodes were angled vertically and positioned on the midclavicular line of the third and the forth intercostal spaces or on the left sternal border of the fourth and fifth intercostal space and when angled at 45° to the right on the left sternal border of the third and the forth intercostal space.ConclusionsExploring areas around the left sternal border of the fourth intercostal space – where the R wave amplitudes are least affected by body positions – is a reliable method for choosing implantation locations for ILRs.
Journal: Journal of Arrhythmia - Volume 31, Issue 3, June 2015, Pages 130–136