کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2963505 1178562 2008 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Late giant coronary aneurysm associated with a fracture of sirolimus eluting stent: A case report
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Late giant coronary aneurysm associated with a fracture of sirolimus eluting stent: A case report
چکیده انگلیسی

SummaryA 73-year-old female underwent percutaneous coronary intervention (PCI) because of stable angina. An elective PCI for the RCA lesion was first performed with deploying sirolimus eluting stents (SES). Three weeks later, PCI was also provided in the residual LAD lesion. Eight months later, she presented with new angina. CAG revealed an in-stent restenosis in the mid LAD and a large eccentric saccular coronary aneurysm (17 mm × 9 mm) at the proximal RCA. Intravascular ultrasound (IVUS) showed absence of stent struts around the orifice of aneurysm, which suggested a fracture of SES stent. The entry of coronary aneurysm was finally sealed with a polytetrafluoroethylene-covered stent. This report documented a rare case of late giant coronary artery aneurysm associated with a fracture of SES.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cardiology - Volume 51, Issue 1, February 2008, Pages 74–79