کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2967781 1178857 2013 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Automatic mode switching of a dual chamber implantable cardioverter–defibrillator induced by a ventricular escape rhythm
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Automatic mode switching of a dual chamber implantable cardioverter–defibrillator induced by a ventricular escape rhythm
چکیده انگلیسی

This report describes a form of group beating induced by a St Jude dual chamber ICD which interpreted a ventricular escape rhythm (with retrograde conduction) as premature ventricular complexes (PVC). These pacemaker-defined PVCs activated the atrial pace-PVC algorithm in 2 steps. 1. The postventricular atrial refractory period (PVARP) was terminated upon detecting a retrograde P wave within its unblanked portion, and 2. An atrial stimulus was released 330 ms after the end of the PVARP. This response resulted in automatic mode switching because the 330 ms interatrial interval was shorter than the atrial tachycardia detection interval.The arrhythmia may be considered to represent an unusual form of pacemaker escape–capture bigeminy.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Electrocardiology - Volume 46, Issue 2, March–April 2013, Pages 136–139
نویسندگان
, , ,