A method to remove CPR artefacts from human ECG using only the recorded ECG
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
A method to remove CPR artefacts from human ECG using only the recorded ECG
چکیده انگلیسی

SummaryAimTo show the possibility of using cardiopulmonary resuscitation (CPR) artefact suppression methods that do not need additional reference signals to model CPR artefacts.Materials and methodsA CPR suppression method based on a Kalman filter was designed. The artefact was modelled using the fundamental frequency of the compressions, estimated from the spectral analysis of the ECG. Artificial mixtures of human shockable rhythms and CPR artefacts were used to design the algorithm that was then tested on samples obtained from real out-of-hospital cardiac arrest episodes.ResultsThe shock/no-shock decision of an automated external defibrillator (AED) was evaluated before and after CPR suppression for 131 shockable and 347 non-shockable samples. The sensitivity improved from 56% (95% CI, 47–64%) to 90% (95% CI, 84–94%). However, the specificity decreased from 91% (95% CI, 87–93%) to 80% (95% CI, 76–84%).ConclusionsCPR artefacts can be suppressed using methods based on the analysis of the ECG alone. The hardware of current AEDs does not need to be replaced, although better artefact suppression methods exist for modified AEDs with additional reference channels.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Resuscitation - Volume 76, Issue 2, February 2008, Pages 271–278