Clinical evaluation of a new functional test for detection of activated protein C resistance (Pefakit® APC-R Factor V Leiden) at two centers in Europe and the USA
چکیده انگلیسی

A new clotting assay, Pefakit® APC-R Factor V Leiden (Pentapharm Ltd., Switzerland), for the detection of an increased resistance of coagulation factor V against degradation by activated protein C, caused mainly by the factor V Leiden mutation, was evaluated in clinical studies at two University Centers in Europe and the US. The performance was compared with the performance of the routinely used predicate device COATEST® APC™ Resistance V (Chromogenix IL, USA). Both tests were run in parallel on a STA®-R analyzer (Diagnostica Stago, France). Samples from subjects undergoing routine laboratory thrombophilia screening were examined, 187 at the Institute of Medical and Chemical Laboratory Diagnostics (IMCLD), University of Vienna, Austria, and 236 at the Duke University Medical Center (DUMC), Durham/Raleigh NC, USA. All samples were analyzed for factor V Leiden mutation and prothrombin 20210 G/A mutation using standard PCR methods. The data show that the Pefakit® APC-R Factor V Leiden assay discriminates very well between healthy controls and carriers of the factor V Leiden mutation, even in patients with lupus anticoagulant or with deficiency in Protein C or Protein S. Furthermore, this new test is able to discriminate well between heterozygous and homozygous carriers of the factor V Leiden mutation. In both studies the Pefakit® assay showed 100% sensitivity and 100% specificity for detection of the factor V Leiden mutation, compared to 93.1% sensitivity and 93.0% specificity for the COATEST® APC™ Resistance V in the IMCLD study and 93.9% sensitivity and 95.6% specificity in the DUMC study. The new test has PCR-like discrimination power which will help to decrease costs by reducing the need for PCR verification of borderline cases.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Thrombosis Research - Volume 119, Issue 1, 2007, Pages 17–26
نویسندگان
, , , ,