کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3120204 1582806 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Forces released during alignment with a preadjusted appliance with different types of elastomeric ligatures
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Forces released during alignment with a preadjusted appliance with different types of elastomeric ligatures
چکیده انگلیسی

Introduction: The purpose of this in-vitro study was to compare the forces generated by new nonconventional elastomeric ligatures (NCEL) and conventional elastomeric ligatures (CEL) during leveling and aligning phases. Methods: The testing model consisted of 5 stainless steel 0.022-in preadjusted brackets for second premolar, first premolar, canine, lateral incisor, and central incisor. The canine bracket was welded to a sliding bar that allowed for different vertical positions. The forces generated by 3 sizes of wires (0.012-, 0.014-, and 0.016-in superelastic nickel-titanium) with the 2 types of elastomeric ligatures at different amounts of upward canine misalignment (1.5, 3, 4.5, and 6 mm) were recorded. Results and Conclusions: Significant differences between CEL and NCEL were found for all tested variables (P <.01) with the exception of the 0.014- and 0.016-in wires at canine misalignment of 1.5 mm. A noticeable amount of force was generated with the NCEL at all 4 canine positions with all 3 wire sizes (from about 50 to about 150 g). With 4.5 mm of canine misalignment or more, the average amount of released force with the CEL was approximately zero.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: American Journal of Orthodontics and Dentofacial Orthopedics - Volume 129, Issue 5, May 2006, Pages 687–690
نویسندگان
, ,