AimTo design and assemble a distraction device with a strain gauge, and establish an animal model for testing it during mandibular distraction osteogenesis.MethodsAn osteotomy was made in the same position at the junction of mandibular angle and body in 12 rabbits. A customised distraction device attached to a strain gauge was used to distract the mandible. To find out which was the better material, both polyurethane and silicone were used as encapsulation material. After distraction, radiographs were taken, and specimens were examined histologically to record bone formation.ResultsDistraction osteogenesis was carried out successfully on all rabbits when the device used polyurethanes as the encapsulation material, it did not transfer the strain signals, but those made with silicone did.ConclusionThe device with silicone used as the encapsulation material can be used for strain-testing during mandibular distraction osteogenesis in a rabbit model.
Journal: British Journal of Oral and Maxillofacial Surgery - Volume 45, Issue 8, December 2007, Pages 633–636